The VMW-800 Series vacuum regulators are designed for applications requiring high precision in a miniature package. Hysteresis free adjustments are made through a non-rising stem. Available with an attractive knob, or plain, allowing installation of OEM knobs. Also available is a flush adjusting shaft, and factory pre-set tamper-resistant models. Units may be base mounted or panel mounted.
Applications:
- Medical Equipment
- Laboratory Instruments
- Pick 'n Place
- Process Control
Specifications
- Regulator Type: Constant Bleed, Relieving
- Wetted Materials: Polysulfone, Buna-N, EPDM, Acetal, Polyurethane, Stainless Steel
- Supply Consumption: < 16 SCIM (260 cc/min) @ 30" (760 mm) Hg supply
- Max Supply: 30” (760 mm) Hg
- Temperature Range: 40° - 150° F (4° - 66° C)
- Effect of Supply Variation: < .03" (.76 mm) per 5" (127 mm) Hg supply change when dead ended
- Repeatability: < ±.05” (1.25 mm) Hg when supply is removed and re-applied
- Sensitivity: 1" (25.4 mm) H2O
- Maximum Flow: 1.0 cu. ft/min. (28 LPM)
- Recommended Filtration: 5 micron
Stocked Configurations Below Can Ship Same Day
Part Number | Description |
---|---|
VMW-800-30-W/K | 30 Hg with Adjusting Knob |